Online sales of Pall Acro Vent Filters. Pall Acro Filters
are available to resellers, University and
Government customers with a starting 10% discount if ordered online or by fax. Bulk discounts on Pall Acro Filters
and all Pall Filters are available upon request.
ORDER PALL ACRO CAPSULE FILTERS ONLINE AND
RECEIVE A 10% DISCOUNT
PALL ACRO FILTERS SPECIFICATIONS
Materials of Construction Filter Media: PTFE on a polypropylene
support Housing: Polypropylene
Effective Filtration Area 19.6 cm2
Dimensions Overall Length: 8.2 cm (3.2 in.) Diameter: 7.3 cm
(2.9 in.)
Inlet/Outlet Connections Stepped hose barbs, 6.4 - 12.7 mm (1/4 -
1/2 in.) diameter 1/8 in. MNPT 9.5 mm (3/8 in.) straight
pipe
Maximum Operating Temperature 130 °C (266
°F)
Maximum Operating Pressure 4.1 bar (410 kPa, 60 psi) at
ambient temperature
Recommended Integrity Test Minimum
Bubble Point – Methanol 0.2 µm: 0.9 bar (90 kPa, 13 psi) 0.45 µm: 0.5
bar (50 kPa, 7 psi) 1 µm: 0.2 bar (20 kPa, 3 psi)
Typical Methanol
Flow Rate mL/min at 1.0 bar (15 psi) 0.2 µm: 350 0.45 µm: 550 1
µm: 700
Typical Air Flow Rates L/min at 0.2 bar (20 kPa, 3
psi) 0.2 µm: 8 0.45 µm: 12 1 µm: 15
Biological
Safety Passes United States Pharmacopeia (USP) Biological Reactivity
Test, In Vivo <88>
Sterilization Provided
non-sterile, individually packaged (except PN 4250). Can be sterilized by
autoclaving at 121 - 123 °C (250 - 253 °F) for 20 min. Integrity should be
verified after each autoclave cycle and before use.